Název: Influence of the Measured Contact Cone Shape Deflections of the Hip Joint Endoprosthesis on the Stress and Reliability Values
Autoři: Fuis, Vladimír ; Hlavoň, Pavel
Typ dokumentu: Příspěvky z konference
Konference/Akce: National Conference with International Participation "Engineering Mechanics 2003", Svratka (CZ), 2003-05-12 / 2003-05-15
Rok: 2003
Jazyk: eng
Abstrakt: The paper presents the results of the computational modelling of the stress states in the ceramic head of the hip joint endoprosthesis. IMS UMPIRE measures 3D shape deflections of the contact cone areas. These measured values were implemented to FEM model. The stress and reliability analyses follow. The values of maximum tensile stresses in the head and head's fracture probability curves create dispersion for random head putting on the stem cone.
Klíčová slova: Hip joint endoprosthesis; manufacturing deflections; stress and reliability analyses
Číslo projektu: CEZ:AV0Z2076919 (CEP), GP101/01/P039 (CEP)
Poskytovatel projektu: GA ČR
Zdrojový dokument: Engineering mechanics 2003, ISBN 80-86246-18-3

Instituce: Ústav termomechaniky AV ČR (web)
Informace o dostupnosti dokumentu: Dokument je dostupný v příslušném ústavu Akademie věd ČR.
Původní záznam: http://hdl.handle.net/11104/0107960

Trvalý odkaz NUŠL: http://www.nusl.cz/ntk/nusl-31563


Záznam je zařazen do těchto sbírek:
Věda a výzkum > AV ČR > Ústav termomechaniky
Konferenční materiály > Příspěvky z konference
 Záznam vytvořen dne 2011-07-01, naposledy upraven 2024-01-26.


Není přiložen dokument
  • Exportovat ve formátu DC, NUŠL, RIS
  • Sdílet