Název: Determination of 5-hydroxyindole-3-acetic acid using Flow Injection Analysis with Electrochemical Detection
Autoři: Makrlíková, Anna ; Matysik, F.-M. ; Navrátil, Tomáš ; Barek, J. ; Vyskočil, V.
Typ dokumentu: Příspěvky z konference
Konference/Akce: Moderní elektrochemické metody /37./, Jetřichovice (CZ), 20170515
Rok: 2017
Jazyk: eng
Abstrakt: Tumour biomarker, 5-hydroxyindole-3-acetic acid, has been determined at screen-printed carbon electrodes using flow injection analysis with amperometric detection. Britton-Robinson buffer (0.04 mol-l-1 , pH 2.0) was used as an optimum carrier solution.\nDependence of the peak current on the analyte concentration was linear in the whole tested\nconcentration range (from 0.5 to 100 mikromol-l-1) with limits of detection 0.033 mikromol-l-1 \n(calculated from peak heighst) and 0.012 mikromol-l-1 (calculated from peak areas).
Klíčová slova: 5-Hydroxyindole-3-acetic acid; Amperometric detection; Flow injection analysis
Číslo projektu: GA17-03868S (CEP)
Poskytovatel projektu: GA ČR
Zdrojový dokument: XXXVII. Moderní elektrochemické metody. Sborník přednášek mezinárodní odborné konference, ISBN 978-80-905221-5-2

Instituce: Ústav fyzikální chemie J. Heyrovského AV ČR (web)
Informace o dostupnosti dokumentu: Dokument je dostupný na vyžádání prostřednictvím repozitáře Akademie věd.
Původní záznam: http://hdl.handle.net/11104/0271996

Trvalý odkaz NUŠL: http://www.nusl.cz/ntk/nusl-317297


Záznam je zařazen do těchto sbírek:
Věda a výzkum > AV ČR > Ústav fyzikální chemie J. Heyrovského
Konferenční materiály > Příspěvky z konference
 Záznam vytvořen dne 2017-06-12, naposledy upraven 2023-12-06.