Original title:
Determination of 5-hydroxyindole-3-acetic acid using Flow Injection Analysis with Electrochemical Detection
Authors:
Makrlíková, Anna ; Matysik, F.-M. ; Navrátil, Tomáš ; Barek, J. ; Vyskočil, V. Document type: Papers Conference/Event: Moderní elektrochemické metody /37./, Jetřichovice (CZ), 20170515
Year:
2017
Language:
eng Abstract:
Tumour biomarker, 5-hydroxyindole-3-acetic acid, has been determined at screen-printed carbon electrodes using flow injection analysis with amperometric detection. Britton-Robinson buffer (0.04 mol-l-1 , pH 2.0) was used as an optimum carrier solution.\nDependence of the peak current on the analyte concentration was linear in the whole tested\nconcentration range (from 0.5 to 100 mikromol-l-1) with limits of detection 0.033 mikromol-l-1 \n(calculated from peak heighst) and 0.012 mikromol-l-1 (calculated from peak areas).
Keywords:
5-Hydroxyindole-3-acetic acid; Amperometric detection; Flow injection analysis Project no.: GA17-03868S (CEP) Funding provider: GA ČR Host item entry: XXXVII. Moderní elektrochemické metody. Sborník přednášek mezinárodní odborné konference, ISBN 978-80-905221-5-2
Institution: J. Heyrovsky Institute of Physical Chemistry AS ČR
(web)
Document availability information: Fulltext is available on demand via the digital repository of the Academy of Sciences. Original record: http://hdl.handle.net/11104/0271996