Original title: Region of interest tomography
Authors: Vopálenský, Michal ; Kumpová, Ivana
Document type: Research reports
Year: 2020
Language: eng
Abstract: During standard tomographic process, the acquisition of tomographic data (projections) is performed with geometric adjustment of the tomographic assembly ensuring that the projection of the entire width of the examined object fits onto the detector during the full object rotation. Because of the limited width of the detector, this approach has a limited maximum achievable resolution for the given specimen diameter. It is possible to make higher magnification, but then the projections do not fit onto the detector, which causes certain problems during the reconstruction, but allows achieving of higher resolution in the always-visible region of the object (the region of interest). The research was focused on evaluation of ROI tomography approach on the TORATOM tomograph.
Keywords: Region-of-Interest tomography; resolution; X-ray computed tomography

Institution: Institute of Theoretical and Applied Mechanics AS ČR (web)
Document availability information: Fulltext is available at the institute of the Academy of Sciences.
Original record: http://hdl.handle.net/11104/0312959

Permalink: http://www.nusl.cz/ntk/nusl-432288


The record appears in these collections:
Research > Institutes ASCR > Institute of Theoretical and Applied Mechanics
Reports > Research reports
 Record created 2020-12-03, last modified 2021-11-24


No fulltext
  • Export as DC, NUŠL, RIS