Název: Region of interest tomography
Autoři: Vopálenský, Michal ; Kumpová, Ivana
Typ dokumentu: Výzkumné zprávy
Rok: 2020
Jazyk: eng
Abstrakt: During standard tomographic process, the acquisition of tomographic data (projections) is performed with geometric adjustment of the tomographic assembly ensuring that the projection of the entire width of the examined object fits onto the detector during the full object rotation. Because of the limited width of the detector, this approach has a limited maximum achievable resolution for the given specimen diameter. It is possible to make higher magnification, but then the projections do not fit onto the detector, which causes certain problems during the reconstruction, but allows achieving of higher resolution in the always-visible region of the object (the region of interest). The research was focused on evaluation of ROI tomography approach on the TORATOM tomograph.
Klíčová slova: Region-of-Interest tomography; resolution; X-ray computed tomography

Instituce: Ústav teoretické a aplikované mechaniky AV ČR (web)
Informace o dostupnosti dokumentu: Dokument je dostupný v příslušném ústavu Akademie věd ČR.
Původní záznam: http://hdl.handle.net/11104/0312959

Trvalý odkaz NUŠL: http://www.nusl.cz/ntk/nusl-432288


Záznam je zařazen do těchto sbírek:
Věda a výzkum > AV ČR > Ústav teoretické a aplikované mechaniky
Zprávy > Výzkumné zprávy
 Záznam vytvořen dne 2020-12-03, naposledy upraven 2021-11-24.


Není přiložen dokument
  • Exportovat ve formátu DC, NUŠL, RIS