Název: Miniaturized boron doped diamond film electrode for neuroblastoma biomarkers determination
Autoři: Hrdlička, Vojtěch ; Barek, J. ; Navrátil, Tomáš ; Taylor, Andrew
Typ dokumentu: Příspěvky z konference
Konference/Akce: INTERNATIONAL STUDENTS CONFERENCE MODERN ANALYTICAL CHEMISTRY /13./, Prague (CZ), 20170921
Rok: 2017
Jazyk: eng
Abstrakt: Cathodically pretreated miniaturized boron doped diamond film electrodewas successfully tested for the purposes of in-situ determination of neuroblastoma biomarkers after hollow-fibre based microextraction. Electrode was tested using K-3[Fe(CN)(6)] and vanillyl-mandelic acid as model substances. Limits of detection and determination are 2.2 and 6.6 mu mol L-1, respectively. Linear range was 2.2-100 mu molL(-1), repeatability 1.34% (n = 10).
Klíčová slova: biomarkers; boron doped diamond film; voltammetry
Číslo projektu: GA17-15319S (CEP)
Poskytovatel projektu: GA ČR
Zdrojový dokument: Proceedings of the 13th International Students Conference ”Modern Analytical Chemistry”, ISBN 978-80-7444-052-6

Instituce: Fyzikální ústav AV ČR (web)
Informace o dostupnosti dokumentu: Dokument je dostupný v příslušném ústavu Akademie věd ČR.
Původní záznam: http://hdl.handle.net/11104/0301585

Trvalý odkaz NUŠL: http://www.nusl.cz/ntk/nusl-407926


Záznam je zařazen do těchto sbírek:
Věda a výzkum > AV ČR > Fyzikální ústav
Konferenční materiály > Příspěvky z konference
 Záznam vytvořen dne 2019-11-25, naposledy upraven 2021-11-24.