Název: Correction of gradient pulse shape distortions in radial MRI
Autoři: Vitouš, Jiří
Typ dokumentu: Příspěvky z konference
Konference/Akce: STUDENT EEICT 2022 /28./, Brno (CZ), 20220426
Rok: 2022
Jazyk: eng
Abstrakt: This paper focuses on the optimization of gradient-pulse shapes in MRI measurement. The main topic investigated in this paper is optimization of slice-selective gradient, where imperfections may produce phase distortion in the resulting image and also signal loss in the acquired signal. A method for correction is proposed based on the Nelder-Mead algorithm followed by coordinate ascent search in the neighborhood of the found solution. The adjustment is evaluated using a simple Fast low angle shot (FLASH) sequence with radial readout. The results show a significant improvement in the Free induction decay (FID) signal magnitude, echo stability, and an improvement in the homogeneity of image phase.
Klíčová slova: adjustment; gradient; MRI; Nelder–Mead; slice
Číslo projektu: LM2018129 (CEP), EF18_046/0016045, GA22-10953S (CEP)
Poskytovatel projektu: GA MŠk, GA MŠk, GA ČR
Zdrojový dokument: Proceedings II of the 28th Conference STUDENT EEICT 2022. Selected Papers, ISBN 978-80-214-6030-0, ISSN 2788-1334

Instituce: Ústav přístrojové techniky AV ČR (web)
Informace o dostupnosti dokumentu: Dokument je dostupný na externích webových stránkách.
Externí umístění souboru: https://www.eeict.cz/eeict_download/archiv/sborniky/EEICT_2022_sbornik_2_v3.pdf
Původní záznam: https://hdl.handle.net/11104/0340112

Trvalý odkaz NUŠL: http://www.nusl.cz/ntk/nusl-521041


Záznam je zařazen do těchto sbírek:
Věda a výzkum > AV ČR > Ústav přístrojové techniky
Konferenční materiály > Příspěvky z konference
 Záznam vytvořen dne 2023-02-26, naposledy upraven 2023-03-28.


Není přiložen dokument
  • Exportovat ve formátu DC, NUŠL, RIS
  • Sdílet