Název:
HPLC-ED/UV with Solid Phase Extraction for the Determination of 5-Hydroxyindole-3-acetic Acid
Autoři:
Němečková-Makrlíková, Anna ; Dejmková, H. ; Navrátil, Tomáš ; Barek, J. ; Vyskočil, V. Typ dokumentu: Příspěvky z konference Konference/Akce: Moderní elektrochemické metody /38./, Jetřichovice (CZ), 20180521
Rok:
2018
Jazyk:
eng
Abstrakt: Determination of 5-HIAA in human urine was successfully performed using HPLC-ED/UV with SPE. HPLC-ED is a very sensitive technique, presented method does not need any sample pre-treatment apart from SPE procedure, which replaced filtration of samples of human urine to avoid problems with clogging a HPLC column. The found native\nconcentrations of 5-HIAA in human urine using HPLC-ED with SPE fully corresponds with published normal concentrations in urine (17.8 - 58.3 micromol l-1). The found native concentrations of 5-HIAA were 18.4 micromol L-1 (calculated from peak height). The presented method is fast and could be used for simple\nscreening of human urine, because HPLC-ED/UV allows simultaneous determination of 5-HIAA, HVA, and VMA as important tumour biomarkers.
Klíčová slova:
electrochemical detection; HPLC; spectrophotometric detection Číslo projektu: GA17-03868S (CEP) Poskytovatel projektu: GA ČR Zdrojový dokument: XXXVIII. Moderní elektrochemické metody. Sborník přednášek, ISBN 978-80-905221-6-9
Instituce: Ústav fyzikální chemie J. Heyrovského AV ČR
(web)
Informace o dostupnosti dokumentu:
Dokument je dostupný na vyžádání prostřednictvím repozitáře Akademie věd. Původní záznam: http://hdl.handle.net/11104/0284335