Název: Helium cryopump for electron optical devices design and preliminary cryogenic testing
Autoři: Dupák, Jan ; Hanzelka, Pavel ; Musilová, Věra
Typ dokumentu: Příspěvky z konference
Konference/Akce: Cryogenics /6./, Praha (CZ), 2000-10-10 / 2000-10-13
Rok: 2000
Jazyk: eng
Abstrakt: A helium cryopump for electron optical devices was designed and preliminary cryogenic parameters of the first pump sample were tested. Low LHe consumption should enable economical operation of the pump without recycling the evaporated He. The pumping cryopanel is cooled by normal LHe down to 4.2 K. A LN2 bath is utilized for the LHe bath thermal shielding. Two additional heat radiation shields cooled by the vapours of the consumed cryoliquids are installed to decrease the LHe and LN2 losses. Three chevron screens placed between the input office and the cryopanel are used to protect the internal parts against direct heat radiation from the parts at ambient temperature. The chevron systems were replaced by low emissivity screens in the preliniary tests. Predicted low evaporation rates of LHe and LN2 were achieved.
Číslo projektu: CEZ:AV0Z2065902 (CEP), GA101/00/0028 (CEP)
Poskytovatel projektu: GA ČR
Zdrojový dokument: Proceedings of the 6th Cryogenics 2000 IIR International Conference - Refrigeration Science and Technology, ISSN 0151-1637
Poznámka: Související webová stránka: mailto:dupak@isibrno.cz

Instituce: Ústav přístrojové techniky AV ČR (web)
Informace o dostupnosti dokumentu: Dokument je dostupný v příslušném ústavu Akademie věd ČR.
Původní záznam: http://hdl.handle.net/11104/0100882

Trvalý odkaz NUŠL: http://www.nusl.cz/ntk/nusl-29479


Záznam je zařazen do těchto sbírek:
Věda a výzkum > AV ČR > Ústav přístrojové techniky
Konferenční materiály > Příspěvky z konference
 Záznam vytvořen dne 2011-07-01, naposledy upraven 2021-11-24.


Není přiložen dokument
  • Exportovat ve formátu DC, NUŠL, RIS
  • Sdílet