Original title: Novel, versatile capillary electrophoresis instrument with laser induced fluorescence for analysis of various lipid peroxidation biomarkers
Authors: Lačná, J. ; Foret, František ; Kubáň, Petr
Document type: Papers
Conference/Event: CECE 2016. International Interdisciplinary Meeting on Bioanalysis /13./, Brno (CZ), 20161017
Year: 2016
Language: eng
Abstract: In this work we have developed a new sensitive method for quantitation of MDA in\nbiological samples by CE-LIF with LOD about 1.7 nM. First part of the work was\ndevoted to the construction of a new CE-LIF system. Then the derivatization process\nof MDA by TBA was optimized. The ability to use this method for the detection of\nMDA in a biological samples, such as EBC and plasma, was also demonstrated. For the\nfirst time it was possible to detect MDA in EBC samples by CE. Other biomarkers are\nnow being studied.
Keywords: biomarkers; CE-LIF; lipids; malondialdehyde
Host item entry: CECE 2016. 13th International Interdisciplinary Meeting on Bioanalysis, ISBN 978-80-904959-4-4

Institution: Institute of Analytical Chemistry AS ČR (web)
Document availability information: Fulltext is available in the digital repository of the Academy of Sciences.