Original title:
Edge Effects in Rotational Viscometry. I. Cylinder - Cylinder (Couette) System, Newtonian Media
Translated title:
Hranové efekty v rotační viskozimetrii. I. Systém válec-válec (Couette), newtonské kapaliny
Authors:
Wein, Ondřej ; Večeř, Marek Document type: Research reports
Year:
2004
Language:
eng Series:
1 Abstract:
[eng][cze] Exact calibration factors for a series of commercially available rotational viscometers are determined by solving the related elliptic boundary-value problem for the rotational flow of Newtonian media under zero Reynolds number (creeping flow approximation). In particular, the edge-effect correction for the Z40 DIN sensor (coaxial cylinders) should be 1.085 instead of incorrect 1.100, given in the related international standard ISO 3219 (Coaxial cylinders for liquids of medium viscosity). The edge-effect correction was determined also for a new KK sensor (coaxial Morse cones) in dependence of the adjustable gap thickness.Přesné kalibrační konstanty u řady komerčně vyráběných viskozimetrů jsou stanoveny řešením odpovídajících okrajových úloh pro rotační tok newtonských kapalin s Reynoldsovým číslem nižším než nula (aproximace toku). Korekce na hranové efekty pro sensor Z40 DIN (koaxiální válce) by však místo nesprávné hodnoty 1.000 (uvedené v normě ISO 3219 (koaxiální válce pro kapaliny se střední viskozitou) měla být 1.085. Korekce hranových efektů byla určena také pro nový KK senzor (koaxiální Morse kužel).
Keywords:
edge correction; newtonian media; rotational viscometry Project no.: CEZ:AV0Z4072921 (CEP), GA104/01/0545 (CEP) Funding provider: GA ČR
Institution: Institute of Chemical Process Fundamentals AS ČR
(web)
Document availability information: Fulltext is available at the institute of the Academy of Sciences. Original record: http://hdl.handle.net/11104/0012019