Original title:
Early Defect Detection of Acetabular Implants
Authors:
Kytýř, Daniel ; Jiroušek, Ondřej ; Zlámal, Petr ; Doktor, Tomáš ; Jandejsek, I. Document type: Papers Conference/Event: Engineering Mechanics 2012 /18./, Svratka (CZ), 2012-05-14 / 2012-05-17
Year:
2012
Language:
eng Abstract:
The paper is focused on possibilities of modern X-ray detectors and micro-focus X-ray source for investigation of early degradation processes of acetabular implants. To simulate the most adverse activity (downstairs walking) a hip joint simulator was developed. The experimental setup was designed for cyclic loading of polyethylene acetabular cup implanted into the human pelvic bone and fixed by commercial polymethyl methacrylate bone cement. To predict the bone degradation numerical analysis of detailed three-dimensional model of the acetabular cup and the cement mantle implanted in a bone block was performed. Using large area flat panel detector and microfocus X-ray source it is possible to investigate micro-damage propagation and detect early defect in the bone-implant interface.
Keywords:
bone-cement interface; computed tomography; crack detection; hip simulator Project no.: GAP105/10/2305 (CEP) Funding provider: GA ČR Host item entry: Engineering Mechanics 2012 - CD ROM, ISBN 978-80-86246-40-6 Note: Související webová stránka: http://www.engmech.cz/
Institution: Institute of Theoretical and Applied Mechanics AS ČR
(web)
Document availability information: Fulltext is available at the institute of the Academy of Sciences. Original record: http://hdl.handle.net/11104/0209666