Original title:
Lab-made Sensors Based on Boron-doped Diamond for Determination of Herbicide Linuron
Authors:
Štěpánková, M. ; Šelešovská, R. ; Janíková, L. ; Vojs, M. ; Marton, M. ; Behúl, M. ; Nováková, Kateřina ; Chýlková, J. Document type: Papers Conference/Event: Moderní elektrochemické metody /35./, Jetřichovice (CZ), 20150518
Year:
2015
Language:
eng Abstract:
Voltammetric behaviour of substituted urea herbicide linuron (LIN) was investigated using two types of boron-doped diamond electrodes (BDDE): commercial BDDE and lab-made sensors based on boron-doped diamond (LM-BDDE). Various methods like cyclic voltammetry (CV), direct current voltammetry (DCV) and differential pulse voltammetry (DPV) were examined. It was found that LIN provides one oxidation peak using all of the tested working electrodes and the highest current response was recorded in the acidic media. Thus, Britton-Robinson (B-R) buffer of pH 2 was used as a supporting electrolyte. Parameters of DPV were optimized and the low limits of detection (LOD) were reached 1.41×10−7 M for commercial boron-doped diamond electrode and 8.81×10−9 M, respectively, which represents the lowest LOD obtained for lab-made boron-doped diamond electrodes.
Keywords:
Boron doped diamond electrode; sensors; voltammetry Host item entry: XXXV. Moderní elektrochemické metody. Sborník přednášek mezinárodní odborné konference, ISBN 978-80-905221-3-8
Institution: J. Heyrovsky Institute of Physical Chemistry AS ČR
(web)
Document availability information: Fulltext is available at the institute of the Academy of Sciences. Original record: http://hdl.handle.net/11104/0246994