Název:
Calibration of germanium temperature sensors of helium cryostat for experimental study of natural turbulent convection
Autoři:
Urban, Pavel ; Hanzelka, Pavel ; Musilová, Věra Typ dokumentu: Příspěvky z konference Konference/Akce: Cryogenics 2010 /11./, Bratislava (SK), 2010-04-26 / 2010-04-29
Rok:
2010
Jazyk:
eng
Abstrakt: Experimental study of Nusselt number on Rayleigh number dependence with using of cryogenic 4He gas requires precise measurement of the temperature differenceT between the top and the bottom plates of the convection cell. A special calibration device was adapted and used for comparative calibration of germanium temperature sensors. This device enables to compare up to four sensors within the temperature range of 4.3 K – 40 K.
Klíčová slova:
calibration device; comparative calibration; natural convection; Rayleigh number; temperature sensors calibration Číslo projektu: CEZ:AV0Z20650511 (CEP), ED0017/01/01, KJB200650902 (CEP) Poskytovatel projektu: GA MŠk, GA AV ČR Zdrojový dokument: The 11th Cryogenics 2010 - IIR International Conference, ISBN 978-2-913149-77-9
Instituce: Ústav přístrojové techniky AV ČR
(web)
Informace o dostupnosti dokumentu:
Dokument je dostupný v příslušném ústavu Akademie věd ČR. Původní záznam: http://hdl.handle.net/11104/0191836