Original title:
Impact Hammer Calibration
Authors:
Bilík, Šimon Document type: Papers
Language:
cze Publisher:
Vysoké učení technické v Brně, Fakulta elektrotechniky a komunikačních technologií Abstract:
This paper describes the construction of the calibration station for the impact hammer Endevco 2302-5 and it explains the influence of the impact mass on the calibration process. Based on the previous bachelor thesis [1] it also explains the origin of the oscillations on the accelerometer output signal described there and it suggests the improvements of the calibration process.
Keywords:
accelerometers; calibration; impact hammer Host item entry: Proceedings of the 25st Conference STUDENT EEICT 2019, ISBN 978-80-214-5735-5
Institution: Brno University of Technology
(web)
Document availability information: Fulltext is available in the Brno University of Technology Digital Library. Original record: http://hdl.handle.net/11012/186677