Original title:
Hip joint ceramic head reliability versus contact cone production inaccuracies
Authors:
Fuis, Vladimír Document type: Papers Conference/Event: Conference of the Czech Society of biomechanics with international participation Biomechanics of man 2000 /8./, Olomouc (CZ), 2000-11-24 / 2000-11-25
Year:
2000
Language:
eng Abstract:
The paper contains the results of the stress and reliability analyses of the total hip joint endoprosthesis ceramic head. The modelled system consists of the two parts: ceramic head and steel stem. The ordering of mentioned system is according the norm ISO 7206-5. The model production inaccuracies of the contact taper of the head and the stem are considered. The head stresses are obtained from the FEM solution and the head reliability is based on the Weibull weakest link theory.
Keywords:
F.E.M. modelling; hip joint endoprosthesis Project no.: 20/7U Funding provider: ÚT AV ČR Host item entry: Biomechanics of man 2000, ISBN 80-244-0193-2
Institution: Institute of Thermomechanics AS ČR
(web)
Document availability information: Fulltext is available at the institute of the Academy of Sciences. Original record: http://hdl.handle.net/11104/0107149