Original title: Result of Extended Type Tests of the liquid Scintillation Spectrometer TriCarb 3170TR/Sl
Authors: Soukup, T. ; Světlík, Ivo
Document type: Papers
Conference/Event: Days of Radiation protection /XXV./, Jáchymov (CZ), 2002-11-27 / 2002-11-29
Year: 2002
Language: eng
Abstract: The liquid scintillation spectrometer TriCarb 3170 TR/SL was installed in laboratories of the NPP Temelín for purpose of monitoring of tritium and some other beta emitting radionuclides in the environment. Tritium was chosen as a testing radionuclide. Stability of measurements in case of samples containing higher activities and of strongly quenched samples with relatively small activities of tritium was the main emphasis in these tests. Results of statistical evaluation confirmed high stability and considerable robustness of the measurements performed by the TriCarb 3170 TR/SL.
Keywords: metrology; quenching; tritium
Project no.: CEZ:AV0Z1048901 (CEP)
Host item entry: Conference Abstracts XXV. Days of radiation protection, ISBN 80-01-02651-5

Institution: Nuclear Physics Institute AS ČR (web)
Document availability information: Fulltext is available at the institute of the Academy of Sciences.
Original record: http://hdl.handle.net/11104/0082044

Permalink: http://www.nusl.cz/ntk/nusl-27795


The record appears in these collections:
Research > Institutes ASCR > Nuclear Physics Institute
Conference materials > Papers
 Record created 2011-07-01, last modified 2024-01-26


No fulltext
  • Export as DC, NUŠL, RIS