Název:
Correction of gradient pulse shape distortions in radial MRI
Autoři:
Vitouš, Jiří Typ dokumentu: Příspěvky z konference
Jazyk:
eng
Nakladatel: Vysoké učení technické v Brně, Fakulta elektrotechniky a komunikačních technologií
Abstrakt:
This paper focuses on the optimization of gradient-pulse shapes in MRI measurement. The main topic investigated in this paper is optimization of slice-selective gradient, where imperfections may produce phase distortion in the resulting image and also signal loss in the acquired signal. A method for correction is proposed based on the Nelder-Mead algorithm followed by greedy search in the neighborhood of the found solution. The adjustment is evaluated using a simple Fast low angle shot (FLASH) sequence with radial readout. The results show a significant improvement in the Free induction decay (FID) signal magnitude, echo stability, and an improvement in the homogeneity of image phase.
Klíčová slova:
MRI, gradient, Nelder-Mead, slice, adjustment Zdrojový dokument: Proceedings II of the 28st Conference STUDENT EEICT 2022: Selected papers, ISBN 978-80-214-6030-0
Instituce: Vysoké učení technické v Brně
(web)
Informace o dostupnosti dokumentu:
Plný text je dostupný v Digitální knihovně VUT. Původní záznam: http://hdl.handle.net/11012/208649