Original title:
Použití pyroelektrického senzoru TGS v pokusném infračerveném analyzátoru plynů
Translated title:
Application of pyroelectric sensor TGS in experimental infrared gas analyzer
Authors:
Zelinka, Jiří Document type: Research reports
Year:
2005
Language:
cze Series:
Výzkumná zpráva, volume: Z 2096/A Abstract:
[cze][cze] The research report descibes the utilization of a triglycine sulfate pyroelectric sensor in an experimental infrared gas analyzer designed in IREE for the control of the carbon dioxide concentration in gaseous mixtures. The report includes details about the construction of the analyzer, describes its function, method of calibration and its medium term stability. Information gained during the development and testing of the analyzer is also included and can be considered in case of future further development of this unit.Výzkumná zpráva popisuje použití pyroelektrického senzoru na bázi síranu triglycinia pro analýzu obsahu kysličníku uhličitého v jednoduchém pokusném infračerveném plynovém analyzátoru. Obsahuje konstrukční údaje a popis činnosti analyzátoru, zjištění jeho kalibrační křivky a střednědobé stability. Uvádí poznatky z pokusných úprav zapojení a z provozování analyzátoru použitelné pro případný další vývoj přístroje.
Keywords:
feeroelectric materials; gas mixtures; pyroelectric detectors Project no.: CEZ:AV0Z20670512 (CEP), IBS2067204 (CEP) Funding provider: GA AV ČR
Institution: Institute of Photonics and Electronics AS ČR
(web)
Document availability information: Fulltext is available at the institute of the Academy of Sciences. Original record: http://hdl.handle.net/11104/0111856