National Repository of Grey Literature 1 records found  Search took 0.00 seconds. 
Determination of indican on carbon paste electrode and on carbon composite electrode
Bergerová, Michaela ; Dejmková, Hana (advisor) ; Schwarzová, Karolina (referee)
This thesis is focused on the application of two working electrodes of graphite suitable for determination of oxidizable tumour biomarkers. It deals with the study of electrochemical behavior of urinary indican. For the determination the technique of differential pulse voltammetry (DPV) was used on carbon paste electrode and film composite paste electrode in Britton-Robinson (BR) buffer. In the first step optimal conditions were determined to find the lowest limit of determination (LQ) of this substance. BR buffer of pH 3 was chosen as the optimum medium for determination on a carbon paste electrode and BR buffer of pH 2 was used as the optimum medium for the determinaton using a film composite electrode. Performance for both electrodes was observed in a concentration range of from 1 · 10-6 to 5 · 10-5 mol·l-1 . The limit of determination for measurements on a carbon paste electrode was 0.7 µmol·l-1 and the film composite electrode was 1.7 µmol·l-1 . Possibility of the acummulation of indican was examined in order to increase sensitivity and decrease of limit of quantification. However, after 5 minute sorption, no increase in response was observed and the accumulation step was not inserted. Developed method was tested on the determination of urinary indican in human urine matrix after solid phase...

Interested in being notified about new results for this query?
Subscribe to the RSS feed.