Název:
Comparison of Results Optained Using Monte Carlo and Ansys Fluent in Analysis Differentially Pumped Chamber
Autoři:
Bílek, Michal ; Hlavatá, Pavla Typ dokumentu: Příspěvky z konference
Jazyk:
eng
Nakladatel: Vysoké učení technické v Brně, Fakulta elektrotechniky a komunikačních technologií
Abstrakt:
The goal of this thesis is the analysis of a gas flow running through the drain channels within the apertures of differentially pumped chamber of the Environmental Scanning Electron Microscope (ESEM). This thesis contains a verification of current simulation results of gas flow within the differentially pumped chamber published by D. Danilatos using Monte Carlo method in comparison with simulation results achieved by using simulation program ANSYS Fluent, which uses the mechanism of continuum for its calculations.
Klíčová slova:
ANYSY Fluent; Environmental Scanning Electron Microscope; ESEM; Monte Carlo Zdrojový dokument: Proceedings of the 22nd Conference STUDENT EEICT 2016, ISBN 978-80-214-5350-0
Instituce: Vysoké učení technické v Brně
(web)
Informace o dostupnosti dokumentu:
Plný text je dostupný v Digitální knihovně VUT. Původní záznam: http://hdl.handle.net/11012/84003