Home > Conference materials > Papers > An Improved Apparatus for Vapor–liquid Equilibria Measurement in Systems Containing Polymer at Higher Temperatures.
Original title:
An Improved Apparatus for Vapor–liquid Equilibria Measurement in Systems Containing Polymer at Higher Temperatures.
Authors:
Pavlíček, Jan ; Bogdanić, Grozdana ; Wichterle, Ivan Document type: Papers Conference/Event: European Symposium on Applied Thermodynamics - ESAT 2017 /29./, Bucharest (RO), 20170518
Year:
2017
Language:
eng Abstract:
The apparatus for total pressure measurement in polymer + solvent systems has been designed recently. In the presented work, the range of functioning of this apparatus was extended to allow investigations at higher temperatures. A direct capacitance pressure sensor (Baratron) was replaced by a pressure transducer, which may be heated up to 200 oC. In addition, the solutions with higher concentrations of polymer can be studied in contrast to the ebulliometric method, which is limited because of problems with physical boiling of viscous mixtures.
Keywords:
polymer; solvent; vapor-liquid equilibrium Project no.: GA15-19542S (CEP) Funding provider: GA ČR Host item entry: Book of Abstracts, ISBN N.
Rights: This work is protected under the Copyright Act No. 121/2000 Coll.
Institution: Institute of Chemical Process Fundamentals AS ČR
(web)
Document availability information: Fulltext is available in the digital repository of the Academy of Sciences. Original record: http://hdl.handle.net/11104/0271750